MedDevice-Catalog Sterilization Validation

Sterilization Validation

All sterilization processes require validation of the efficacy and reproducibility of the process. Depending on the type of sterilization, this may be accomplished by partial, sub-lethal, or repetitive processing, using representative product and/or biological indicators. AppTec offers a full range of services in this area, from testing alone to full management of the validation.

Reusable Medical Device Validation

The FDA expects manufacturers to validate all reuse instructions including cleaning and disinfection procedures, cycle parameters, and aeration times, if applicable. AppTec offers a comprehensive program for evaluation of cleaning and sterilization processes for reusable medical devices.

 

Your AppTec Account Manager can provide you with initial information regarding
any of these testing programs. Also working closely with you will be a
highly trained and knowledgeable technical expert

 

           In this section

EO Half Cycle Sterilization Validation                          Back to Top
For validation of ethylene oxide (EO) sterilization, certain steps must be followed as outlined in ISO and AAMI standards. As part of the performance qualification, a microbiological challenge must be performed to demonstrate the adequacy of the process to achieve the desired sterility assurance level (SAL). One of the most utilized methods is the half-cycle (overkill) method, which uses a biological indicator (BI) challenge, typically 106 spores of Bacillus subtilis. Complete EO validation studies can be designed for a particular product and process, and all aspects of the studies follow the requirements of ISO and AAMI standards.

 

NOTE: Samples for EO Half Cycle Sterilization Validation testing should be shipped to AppTec’s St. Paul facility.

 

ESTIMATED TOTAL TESTING TIME: 4 - 6 weeks

1601000

Devices are selected at random for the bioburden recovery validation. They must be representative of the Worst Case Representative Family Designs being tested, but need not be identified with the production batches being used for validation. 

 

SAMPLE REQUIREMENTS 5-10 products

 

For more information on these tests, see Page A-3.

Bioburden Recovery Efficiency - Repetitive Recovery Method

OR


1602000

Bioburden Recovery Efficiency - Spore Inoculation Method


1603010Non-sterile devices are tested for bioburden.

 

SAMPLE REQUIREMENTS Minimum 10 products

Aerobic Bioburden Panel


1203000

Twenty (20) samples will contain the microbial challenge system (process challenge device) inoculated in the device's most difficult area to sterilize. The BI seeded samples can be simulated products if they are wholly representative of the actual finished product. 

 

SAMPLE REQUIREMENTS 80 products (four sets of 20)

 

[See table on next page for sample size if chamber is larger than two (2) pallets.]

Inoculated Biological Indicators

 


USP Product Sterility Test

Twenty (20) product samples are removed from the load for sterility testing. (A minimum of 40 extra samples should be retained in the event a retest is required.)

 

SAMPLE REQUIREMENTS 20 products

1220010
Extra Small [≤ 100 ml of media]
1220000
Small [101 to 200 ml of media]
1226000
Medium [201 to 400 ml of media]
1227000
Large [401 to 600 ml of media]
1228000
Extra Large [601 to 1200 ml of media]


190105

Sample device or material is tested for growth inhibition using the current USP organisms for Soybean-Casein Digest Medium (SCDM) and Fluid Thioglycollate Medium (FTM). (Additional organisms available upon request.) 

 

SAMPLE REQUIREMENTS 6 products

Bacteriostasis/Fungistasis

 


1201000

A minimum of five samples per pallet, but at least 10 samples per cycle will contain inoculated carriers (spore strips). These carriers will be tested for sterility according to client specifications and/or the USP. 

 

SAMPLE REQUIREMENTS Set of 10
 

[See table below for sample size if chamber is larger than two (2) pallets.]

Biological Indicators

 


195000

Samples from the terminal sterilization cycle will be tested for ethylene oxide residuals at incremental time periods. Suggested time periods are: immediately after sterilization, one day post-sterilization, three days post-sterilization, five days post-sterilization, and seven days post-sterilization. Additional time increments may be required and will be identified in the final report. 

 

SAMPLE REQUIREMENTS Varies

EO Residual Panel - 
Water Extraction

OR


194500

EO Residual Panel - 
Headspace Extraction
ISO 1135 Performance Qualifications:  Sensor and BI Sample Requirements
15 
pallet
11 
pallet

pallet

pallet

pallet

pallet
30 
cu. ft.
25 
cu. ft.
Load Dimensions
Width48"48"48"48"42"42"21"23"
Length48"48"48"48"48"48"60"51"
Height78"78"78"75"72"72"30"35"
Cubic Feet156011449367002521682224
performance qualifications
Commissioning/Physical
Preconditioning: TC/RH Sensor40/1929/1425/1219/99/47/35/25/2
Empty Chamber Temp Profile:
TC Sensor 
2020202012101010
Conditioning: TC/RH Sensor20/1920/1420/1220/912/410/310/210/2

Microbiological

Biological Indicators (BI)5442373024202020

Routine Process

Biological Indicators (BI)2721191512101010

TC = Thermocouples      RH = Relative Humidity

f

Radiation Sterilization Validation                                 Back to Top

For validation of radiation (gamma, electron beam or x-ray) sterilization, certain steps must be followed as outlined in ISO and AAMI standards. As part of the performance qualification, a dose-setting or dose substantiation study must be performed to demonstrate the adequacy of the minimum dose to achieve the desired sterility assurance level (SAL). Several methods are available for validation of the minimum SAL dose, and the choice of method is dependent on a number of variables. Complete radiation validation studies can be designed for a particular product and process, and all aspects of the studies follow the requirements of ISO and AAMI standards.

AppTec offers three levels of service programs for sterilization validations,
as well as two levels of service programs for dose audits.
For descriptions of these programs, see below.

 

NOTE: Samples for radiation sterilization validation testing should be shipped to AppTec’s Atlanta facility.

test namesamples requiredtest description
AAMI/ISO 11137 Method 1 Validation
total turnaround time: 4-6 weeks
Total Bioburden Panel10 products from first batchsee section A (Bioburden/Microbial Assays)
Aerobic Bioburden Panel10 products each from 2nd & 3rd batchessee section A (Bioburden/Microbial Assays)
Bioburden Recovery Efficiency Test3-5 products from any batchsee section A (Bioburden/Microbial Assays)
Bacteriostasis / Fungistasis – One Medium3 products from any batchsee section A (Bioburden/Microbial Assays)
Product Test of Sterility100 samples from one batchsee section G
(Sterility Testing)
Dose Audit
Aerobic Bioburden Panel10 products from one batchsee section A (Bioburden/Microbial Assays)
Product Test of Sterility100* samples from one batchsee section G
(Sterility Testing)
m
AAMI/ISO 11137 Method 2 Validation
total turnaround time: 6-7 weeks
The requirements of Method 2 are not outlined here due to the complexity of the sampling scheme. 
Please contact the laboratory for more information.
Dose Audit
Aerobic Bioburden Panel10 products from one batchsee section A (Bioburden/Microbial Assays)
Product Test of Sterility100* samples from one batchsee section G
(Sterility Testing)
*Reduced number of samples may apply. Contact laboratory.
m
m
test namesamples requiredtest description
AAMI TIR 27 (25 kGy) Validation
total turnaround time: 4-6 weeks
Total Bioburden Panel10 products from one batchsee section A (Bioburden/Microbial Assays)
Aerobic Bioburden Panel10 products each from 2nd & 3rd batchessee section A (Bioburden/Microbial Assays)
Bioburden Recovery Efficiency Test5-10 products from one batchsee section A (Bioburden/Microbial Assays)
Bacteriostasis / Fungistasis – One Medium3 products from any batchsee section A (Bioburden/Microbial Assays)
Product Test of Sterility10 samples from one batchsee section G
(Sterility Testing)
Quarterly Dose Audit
Aerobic Bioburden Panel10 products from one batchsee section A (Bioburden/Microbial Assays)
Product Test of Sterility10 samples from one batchsee section G
(Sterility Testing)
m
ISO/TR 13409 (25 kGy) Validation
total turnaround time: 4-6 weeks
Total Bioburden Panel10 products from one batchsee section A (Bioburden/Microbial Assays)
Bioburden Recovery Efficiency Test5-10 products from one batchsee section A (Bioburden/Microbial Assays)
Bacteriostasis / Fungistasis – One Medium3 products from any batchsee section A (Bioburden/Microbial Assays)
Product Test of Sterility10 samples from one batchsee section G
(Sterility Testing)
Quarterly Dose Audit
Aerobic Bioburden Panel10 products from one batchsee section A (Bioburden/Microbial Assays)
Product Test of Sterility10 samples from one batchsee section G
(Sterility Testing)
m
AAMI/ISO 15844 Single-Batch Validation
total turnaround time: 4-6 weeks
Total Bioburden Panel10 products from batchsee section A (Bioburden/Microbial Assays)
Bioburden Recovery Efficiency Test5-10 products from batchsee section A (Bioburden/Microbial Assays)
Bacteriostasis / Fungistasis – One Medium3 products from batchsee section A (Bioburden/Microbial Assays)
Product Test of Sterility100* samples from batchsee section G
(Sterility Testing)

Radiation Sterilization Validation Programs             Back to Top  

To help clients get their products through the validation process, AppTec offer three levels of service options for validation studies and two levels of service options for dose audits – each with varying degrees of involvement.

Your choice of service levels would depend on how much of your company’s time, manpower and expertise you want to commit to your validation program. With the Level 1 option, we provide only the testing services. You schedule and manage all aspects of the irradiation services and you pull together all the documentation to present an organized study. With Level 2, in addition to the testing, we handle all the irradiation services. However, you are still responsible for producing the final documentation. With Level 3, we take care of everything. You simply choose the validation method, give us your product samples, and 4 to 6 weeks later we give you a completed validation and finished manual.

 

These programs are conducted by AppTec’s Atlanta facility.
Contact your Account Manager for more information.

 

V a l i d a t i o n   P r o g r a m s

LEVEL 1


Testing Services

We perform the full panel of the specific tests required by the particular validation method you have chosen –– AAMI/ISO Method 1, Method 2, 
the 25 kGy Method, or the Single-Batch Method.


[See above charts]

LEVEL 2

 


Testing Services + Irradiation Services
We perform all the required testing plus we provide the verification irradiation services –– we compute the dose calculations, coordinate all interaction with the sterilizer, ensure that samples are irradiated at the verification dose, and provide the confirmation of irradiation.

LEVEL 3

 

Total Management of the Validation Study
We perform all the required testing, provide verification irradiation services and produce a manual that includes the protocol, test reports and supporting documents as attachments to protocol, the official AAMI / ISO standards, irradiation documentation (instructions, dose certificates, etc.), and a final report summarizing the study.


Most programs that are validated for radiation sterilization require periodic dose audits.  
AppTec offers two levels of service options for performing these audits.

D o s e   A u d i t   P r o g r a m s

LEVEL 1


Testing Services

We perform all the tests required for a particular dose audit.

LEVEL 2

 


Testing Services + Irradiation Services
We perform all the required testing plus we provide dose audit irradiation services and a dose audit summary. The summary contains the steps of the dose audit, provides a conclusion, and includes reports and supporting documents.

D o s e   A u d i t   R e m i n d e r   P r o g r a m
The dose audit requirements for Method 1 and Method 2 validations now allow for reduction in the number of verification samples as well as a reduction in the frequency, based on certain criteria. 
AppTec offers a dose audit reminder program that informs clients of their options and issues reminders when dose audits are due for each product or family of products.

 

T e s t   C o d e s

validation method codes                             dose audit codes

LEVEL 1

LEVEL 2

LEVEL 3

LEVEL 1

LEVEL 2

AAMI/ISO 11137 Method 1

200100

200110

200120

 

200130

200140

AAMI/ISO 11137 Method 2

200300

200400

200600

 

200800

200900

AAMI TIR 27 (25 kGy) Method

200101

200102

200121

 

200131

200141

ISO13409 (25 kGy) Method

200150

200160

200170

 

200180

200190

AAMI/ISO 15844 Single-Batch Method

200200

200500

200700

 

N/A

N/A

 

NOTE: These test codes are for prepackaged/bundled programs. Clients may select services within any level.

f
Reusable Medical Device Validation                                Back to Top
When reusable medical devices are cleaned and sterilized in a health care facility, manufacturers are responsible for providing their customers with complete and comprehensive written instructions for handling, cleaning, disinfection and sterilization. The FDA expects manufacturers to validate all reuse instructions including cleaning and disinfection procedures, cycle parameters, and aeration times, if applicable.

 

AppTec offers a comprehensive program for evaluation of cleaning and sterilization processes for reusable medical devices. Testing follows the guidelines outlined in the AAMI Technical Information Report No. 12 ("Designing, Testing & Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: A Guide for Device Manufacturers"). This program assists the manufacturer in meeting the requirements of the FDA Reviewer Guidance: "Labeling Reusable Devices for Reprocessing in a Health Care Facility."

All samples for reusable medical device testing should be shipped to AppTec’s St. Paul facility.

 

Protocol Development

A custom protocol is written for each study, tailored specifically to the device and the manufacturer’s instructions for reuse. AppTec’s scientific staff assists clients in designing cleaning processes and developing protocols.

Cleaning Efficacy Studies

Manufacturers must verify the efficacy of their recommended cleaning processes. Following the manufacturer’s cleaning instructions, this study tests those processes using simulated soil inoculated with bacterial spores.

Sterilization Efficacy Studies

Manufacturers must provide health care facilities with detailed sterilization instructions for their particular medical device. Sterilization parameters are tested to determine capability of producing a sterility assurance level of at least 10-6. Studies are available to evaluating the following sterilization processes:

 

• Liquid chemical sterilants                                          • Pre-vacuum, steam (121-123°C and 132-135°C)
• SterradTM (H2O2)                                                            • Gravity, steam (121-123°C and 132-135°C)
• Ethylene Oxide (EO)

Support for Functionality Studies 

These studies involve exposure to multiple cleaning and or sterilization cycles as part of the functionality studies required to determine the useful life of a device.

For more information on reusable device validation testing, contact your Account Manager.

Back to Top